• Single Double Blood Pressure Monitoring Set
  • Single Double Blood Pressure Monitoring Set
  • Single Double Blood Pressure Monitoring Set
  • Single Double Blood Pressure Monitoring Set
  • Single Double Blood Pressure Monitoring Set
  • Single Double Blood Pressure Monitoring Set

Single Double Blood Pressure Monitoring Set

Type: Catheter
Material: Plastic
Ethylene Oxide Sterilization: Ethylene Oxide Sterilization
Quality Guarantee Period: Two Years
Group: Adult
Logo Printing: With Logo Printing
Customization:
Gold Member Since 2014

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Model NO.
ZG-CZWT38
Delivery
Based on Order Quantity, Commomly Within 30days
Transport Package
Carton
Specification
CE
Trademark
Surecare
Origin
China
HS Code
9018
Production Capacity
10000 PCS Per Week

Product Description

medical Utah disposable Invasive Blood Pressure IBP Transducer disposable pressure transducer ibp

Main technical parameters and working environment requirements:

Disposable Pressure Transducer 
Barometric pressure: 70~106Kpa
Relative humidity: 10~90% (non-condensing)
Operating pressure range: -50 ~ + 300 mmHg 
Sensitivity: 5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis: full-scale reading±1.5% 
Input impedance: 1200Ω ~3200Ω 
Output impedance: 300±5% 
Zero pressure offset: -20mmHg~+20mmHg 
Thermal offset shift: ≤±0.3mmHg/°C 
Offset drift: After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift: ≤±0.1%/°C 
Frequency response: Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand: Defibrillation depends on the final connection of the equipment. 
Leakage current: Leakage current is dependent on the final connection of the equipment 
Overpressure load: -400~+4000mmHg 
Shock resistance : Withstand falling from one meter for three times 
Light sensitivity: When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage. 
Contact with the human body: ≤168h 
Storage and transport environment requirements: - 20 ° C~ + 60 ° C, <90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold. 
Intra-system velocity 
Flow 2-5mL/h 
Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions. 
Product shelf-life: 3 years.
Properties
Medical Materials & Accessories
Product Name
Disposable IBP Transducer 

Single Double Blood Pressure Monitoring SetSingle Double Blood Pressure Monitoring SetSingle Double Blood Pressure Monitoring SetSingle Double Blood Pressure Monitoring SetSingle Double Blood Pressure Monitoring Set

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now