Product Description
Features
lDesigned for multiple drug injection application (up to 8 pumps)
Only one power cord for multi-channel pumps
Innovative Link Mode offers continuous injection to improve safety for patients
Drug library with all the most used drugs
History record review and last configuration memory
Anti-bolus system to reduce significantly bolus after occlusion release
Change the flow rate while operating
Manual bolus and automatic bolus
Wireless data transmission with infusion supervision system and bedside workstation(optional)
Syringe Pump 2-channel Syringe Pump 4-channel Syringe Pump 6-channel Syringe Pump 8-channel Syringe Pump
Syringe Size 5, 10, 20, 30, 50 ml
Operating Mode 1) Flow rate mode, rate range: 0.1 1500.0 ml/h, maximum rate value depends on different syringes
2) Time mode, time range: 1-2000min, Preset Volume: 0.1-999.9ml
3) Body weight mode, weight range: 0.1 300.0kg, drug limit: 0.1-999.9mg, drug unit: mg/g, dose limit: 0.1 1500.0, dose unit: ug/kg/min or mg/kg/h, preset volume range: 0.1 999.9ml
4) Link mode
Flow Rate Range 5 ml syringe: 0.1 ml/h 100 ml/h
10 ml syringe: 0.1 ml/h 200 ml/h
20 ml syringe: 0.1 ml/h 400 ml/h
30 ml syringe: 0.1 ml/h 600 ml/h
50 ml syringe: 0.1 ml/h 1500 ml/h
Increment 0.1ml/h
Accumulated Injection Volume 0.1 999.9 ml
KVO Rate 0.1-5.0ml/h adjustable,1ml/h as default
Drug library 40 drugs
History record 1500 records (Operation status, Infusion Information, Alarms)
Saved syringe brands 5 brands
Bolus Rate 5 ml syringe: 100 ml/h; 10 ml syringe: 200 ml/h; 20 ml syringe: 400 ml/h;
30 ml syringe: 400 600 ml/h adjustable; 50 ml syringe: 600 1500 ml/h adjustable
Accuracy ± 5%
Display
Screen Independent 3.0 inch color TFT screen for each pump channel
Information Operation Mode, Flow Rate, Accumulated Injection Volume, Syringe Size, Battery Capacity, Bed No., AC power indicator, Battery charge indicator, Check the Syringe, etc.
Status Indicator Stop, Run, Bolus, KVO
Alarm Type
Finish, Near, Syringe Empty, Syringe Disengaged, Occlusion, System Error, Battery Empty, Low Battery, Reminder to Operate, No AC Power, Memory Error, Motor Error
Occlusion Alarm Pressure Range: 3 levels: 70kPa, 100kPa, 130kPa (±30kPa), Level Low as default
Battery
Type Lithium polymer battery
Life Up to 2 hours continuous operation@5 ml/h flow rate
Environment
Temperature 5 - 40 c for operating; -40 - 55 c for storage
Air Pressure 86 - 106 kPa for operating; 50 - 106 kPa for storage
Humidity 20 % - 80 % for operating; 10 % - 95 % for storage