Customization: | Available |
---|---|
Type: | Gynaecology |
Function: | Biological Regulatio |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Brief introduction: Air mode and Baby mode controlled by micro-computer; Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance; Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately; Failure alarm indication; Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it; Natural air flow humidity; The height of whole unit can be adjusted; With Oxygen cylinder and Oxygen supply configuration; Observation lamp.
Standard Configuration: Main body (including the hood,controller,bassinet,internal battery,and observation lamp),oxygen supply system,skin temperature sensor,I.V.Pole,mattress,adjustable stand)
Optional Configuration:>37°C override temperature mode
Specification: AC Power supply:AC220V-230V/50HZ
DC Power supply:DC12V/10A or 24V/6A
Input Power:400VA
Mode of control: Air mode and Baby mode controlled by micro-computer
Air temperature control range:25°C-37°C (Override mode):37-38°C
Baby temperature control range: 34°C-37°C (Override mode):37-37.5°C
Skin temperature sensor accuracy: ≤0.3°C
Temperature variability : ≤1.0°C
Temperature uniformity : ≤1.5°C(mattress is on the horizontal position)
Noise inside hood : ≤55dB(A) (Environment noise ≤45Db(A) )
Failure alarm :over-temperature alarm, temperature diviation alarm,sensor alarm,motor fan failed alarm,power failure alarm and so on.
Operating at temperature equilibrium :90 min (Set Temp 36°C, Ambient 15°C)
Environmental temperature: Operating range: 10°C-30°C (Set temperature should be higher 3°C than the environment temperature)
Ambient air movement rate : ≤1.0m/s
Package: Package : each unit is packed as two carton
Size(Main body): 109cm x 65.5cm x 76cm (Cabinet): 138cm x 61cm x 40cm
Gross weight(Main body): 80kg (Cabinet): 53kg |
||||||||||||