Specifications:
1. US
Transducer 9 crystals, wide wave beam
Working mode: Pulsed ultrasound Doppler
Ultrasonic frequency: 1.0MHz± 10%
Measuring range: 50-240 BPM
Alarm range: Up: 55-240 BPM Adjustable continuously within range
Low: 50-235 BPM Adjustable continuously within range
2. TOCO
Display range: 0-100units (non-linear error: ≤ ± 10%)
3. PC PORT
Edition USB 1.1 or USB 2.0; RS-232
4. THERMAL RECORDER
Paper width: 112 ± 2mm effective record width: 105 ± 2mm; Paper length: 15m
Recorder density: US: 27 BPM/ CM; UC: 40 BPM / CM
Printing speed (adjustable): 1 cm/min, 2 cm/min, 3 cm/min
5. POWER and BATTERY
Battery: 16.8V Ni-MH rechargeable batteries group; AC 220 /110 V; Consume power: ≤ 60 W
Continuous working time in static condition: ≥ 1 hours
6. ECG and HR
Measuring range: 30 BPM -350 BPM
Cardiac electric noise level: ≤ 30 μ V
Cardiac electric frequency response:
Filtering mode: 1-25 MHz
Non-filtering mode: 0.05-100 MHz
Cardiac electric Input loop current: ≤ 0.1 μ A
7. NIBP
Measuring range:
A) SBP: (8.034.0) kPa or (60255) mmHg
B) DBP: (4.026.0) kPa or (30195) mmHg
C) MAP: (5.328.7) kPa or (40215) mmHg
Automatic measuring interval: 3 min-99 min adjustable continuously within range
Interval error: ≤ ± 10 s
8. RR
Measuring range: 0 BPM-100 BPM
9. TEMP
Channels: 1
TEMP probe: Surface probe 1pcs
Measuring range: 28-45
Response time: ≤ 30s
10. SpO2
Measuring range: 0-99%
11. PR
Measuring range: 30-250 BPM
12. ENVIROANMENT REQUIREMENT
Working Temperature: +5 -40 Humidity: ≤ 80% Atmospheric pressure: 86 kPa106 kPa
13. ACCESSORIES
2 x US probe, 1 pcs TOCO probe, 1 pcs RS-232 serial port line, 1 pcs probe bandage, 1 pcs SpO2 probe, 1 pcs TEMP probe, 1 pcs electrode probe, 1 pcs power cable, 1 pcs power switch, 1 pcs BP cuff, event marker, 2 pcs printing paper, 250ml gel, , warranty card, operation manual, service manual