medical Utah disposable Invasive Blood Pressure IBP Transducer disposable pressure transducer ibp
Main technical parameters and working environment requirements:
Disposable Pressure Transducer
Barometric pressure: 70~106Kpa
Relative humidity: 10~90% (non-condensing)
Operating pressure range: -50 ~ + 300 mmHg
Sensitivity: 5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis: full-scale reading±1.5%
Input impedance: 1200Ω ~3200Ω
Output impedance: 300±5%
Zero pressure offset: -20mmHg~+20mmHg
Thermal offset shift: ≤±0.3mmHg/°C
Offset drift: After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift: ≤±0.1%/°C
Frequency response: Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand: Defibrillation depends on the final connection of the equipment.
Leakage current: Leakage current is dependent on the final connection of the equipment
Overpressure load: -400~+4000mmHg
Shock resistance : Withstand falling from one meter for three times
Light sensitivity: When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.
Contact with the human body: ≤168h
Storage and transport environment requirements: - 20 ° C~ + 60 ° C, <90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
Intra-system velocity
Flow 2-5mL/h
Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
Product shelf-life: 3 years.